Portable blackbody calibration source, 50℃ to 500℃, 0.15%rdg stability, 120mm diameter plate, 0.95 emissivity, with 6.35mm diameter test hole of PRT
BL675A FLAT FLATE BLACKBODY CALIBRATION SOURCE
The BL675A is a portable blackbody calibration source utilizing a digital indicating temperature ambient 50℃to 500℃. A precision platinum RTD temperature sensor is embedded in the blackbody emitter, providing high accuracy and repeatability. The temperature controller uses the industry standard PID algorithms to control the emitter temperature to within 0.1℃of the set points.
The black body emitter mechanism uses a resistive heating device that provides a long life, short stabilization times, and stable temperature control.
FEATURE
120 mm big aperture diameter, large area naturally minimizes source effect error
Temperature range 50 to 500℃
0.10℃uniformity at 50℃set point
Build-in temperature correction function
Robust design, easy to carry
External 8mm reference sensor input
TYPICAL APPLICATION
Infrared temperature sensors
Infrared thermal imaging systems
Spectrographic analyzers
Spectral radiometers
Heat flux meter
SPECIFICATION
Temperature range
50 to 500℃
Aperture diameter
120 mm
Heated emitter shape
Flat Plate
Emissivity
> 0.95
Display resolution
0.1℃
Control stability (10min)
±0.10℃@ 50℃; ±0.30℃@ 300℃; ±0.40℃@ 500℃
Uniformity (2/3 area)
±0.10℃@ 50℃; ±0.60℃@ 300℃; ±1.00℃@ 500℃
Display accuracy
±0.5°C@ 100°C; ±1°C@ 300°C; ±1.50°C@ 500°C
Control method
Digital PID controller
Temperature sensor
Precision Platinum RTD
Power supply
220V AC / 500WA (110V is optional)
Interface
RS485
Heat up time
60 minutes from ambient to 500℃
Calibration band
8-14um spectral band
Dimension
240mm x 215mm x 375mm
Net weight
10 kg approx.
REFERENCE STANDARD
IEC/TS 62492-1-2008:Industrial process control devices - Radiation thermometers. Part 1: Technical data for radiation thermometers
ASTM E2847-2013: Standard practice for calibration and accuracy verification of wideband infrared thermometers
ASTM E1256-2011a: Standard Test Methods for radiation thermometers (Single waveband type)
OIML D24 (1996): Total radiation pyrometers
JJG 856-2015: Verification regulation of radiation thermometers
SCOPE OF DELIVERY
BL675A Blackbody calibration source
Power cord (need to confirm the plug type in advance)