Infrared thermometer calibrator

Model No.︰

BL675A

Share on:

Product Description

Portable blackbody calibration source, 50℃ to 500℃, 0.15%rdg stability, 120mm diameter plate, 0.95 emissivity, with 6.35mm diameter test hole of PRT

 

BL675A FLAT FLATE BLACKBODY CALIBRATION SOURCE

 

The BL675A is a portable blackbody calibration source utilizing a digital indicating temperature ambient 50 to 500. A precision platinum RTD temperature sensor is embedded in the blackbody emitter, providing high accuracy and repeatability. The temperature controller uses the industry standard PID algorithms to control the emitter temperature to within 0.1 of the set points.

The black body emitter mechanism uses a resistive heating device that provides a long life, short stabilization times, and stable temperature control.

FEATURE

  • 120 mm big aperture diameter, large area naturally minimizes source effect error
  • Temperature range 50 to 500
  • 0.10 uniformity at 50set point
  • Build-in temperature correction function
  • Robust design, easy to carry
  • External 8mm reference sensor input

TYPICAL APPLICATION

  • Infrared temperature sensors
  • Infrared thermal imaging systems
  • Spectrographic analyzers
  • Spectral radiometers
  • Heat flux meter

SPECIFICATION

Temperature range

50 to 500

Aperture diameter

120 mm

Heated emitter shape

Flat Plate

Emissivity

> 0.95

Display resolution

0.1

Control stability (10min)

±0.10@ 50; ±0.30@ 300; ±0.40@ 500

Uniformity (2/3 area)

±0.10@ 50; ±0.60@ 300; ±1.00@ 500

Display accuracy

±0.5°C@ 100°C; ±1°C@ 300°C; ±1.50°C@ 500°C

Control method

Digital PID controller

Temperature sensor

Precision Platinum RTD

Power supply

220V AC / 500WA (110V is optional)

Interface

RS485

Heat up time

60 minutes from ambient to 500

Calibration band

8-14um spectral band

Dimension

240mm x 215mm x 375mm

Net weight

10 kg approx.

REFERENCE STANDARD

  • IEC/TS 62492-1-2008Industrial process control devices - Radiation thermometers. Part 1: Technical data for radiation thermometers
  • ASTM E2847-2013: Standard practice for calibration and accuracy verification of wideband infrared thermometers
  • ASTM E1256-2011a: Standard Test Methods for radiation thermometers (Single waveband type)
  • OIML D24 (1996): Total radiation pyrometers
  • JJG 856-2015: Verification regulation of radiation thermometers

SCOPE OF DELIVERY

  • BL675A Blackbody calibration source
  • Power cord (need to confirm the plug type in advance)
  • Spare fuse (1A for 220VAC, 2A for 110VAC)

OPTIONAL ACCESSORIES

  • Accredited calibration certificate (P/N: 9000-675)
  • USB into RS232 adapter (P/N: 1622)

ORDERING MODEL NUMBER

  • BL675A-110V
  • BL675A-220V


Product Image