Dual Cavities Blackbody calibration source, -15℃ to 500℃, 0.1℃ stability, 65mm and 50mm diameter tube, 0.995 emissivity, with 6.35mm diameter test hole of PRT, 220V AC
BL675D DUAL BLACKBODY CALIBRATION SOURCE
The BL675D Dual Cavity Blackbody Calibration Source is easy to use, and makes calibration of infrared pyrometers with multiple temperature ranges faster than ever before by combining a hot and cold side in one unit.
The calibrators are simple to set and control. Merely input the desired Black Body temperature on the convenient front panel control buttons, let the unit reach the set temperature, wait a few moments to ensure stability, aim, and shoot.The radiated energy from the Black Body is measured by operator's IR thermometer. Compare the IR thermometer reading to the display on the controller and record the difference.
FEATURE
Ideal for simple 2-point calibrations.
Ultra-High uniformity and stability
0.995 emissivity
Fast Heat Up & Cool Down Times
Built-in temperature correction function
Rugged & Lightweight
New Feature: Reference sensor input
APPLICATION
Infrared temperature sensors
Infrared thermal imaging systems
Spectrographic analyzers
Spectral radiometers
Heat flux meter
SPECIFICATION
Low Range
High Range
Temperature range
-15°C ~ 100℃
50°C~ 500°C
Display resolution
0.01°C
0.1°C
Aperture size
65mm
50mm
Distance to target
130mm
160mm
Emissivity
Better than 0.995
Stability (10 min)
0.1°C or 0.1%rd (whichever is greater)
Uniformity
0.15°C or 0.15%rd (whichever is greater)
Heat cavity shape
End closed tube
Cool down time
30 minutes from ambient to -15℃
Heat up time
60 minutes from ambient to 500℃
Control method
Digital PID controller with readout
Temperature sensor
Precision platinum RTD
Power supply
220VAC, 850WA (110VAC is optional)
Interface
RS485
Dimension
435mm(D) x 390mm(W) x 200mm(H)
Net weight
15 kg approx.
REFERENCE STANDARD
IEC/TS 62492-1-2008:Industrial process control devices - Radiation thermometers. Part 1: Technical data for radiation thermometers
ASTM E2847-2013: Standard practice for calibration and accuracy verification of wideband infrared thermometers
ASTM E1256-2011a: Standard Test Methods for radiation thermometers (Single waveband type)
OIML D24 (1996):Total radiation pyrometers
JJG 856-2015: Verification regulation of radiation thermometers
SCOPE OF DELIVERY
BL675D Dual blackbody calibration source
Power code(need to confirm the plug type in advance)